| Size |
19.3"H x 12.3"W x 7.1"D |
| Weight |
17.9 lbs with Power Cartridge |
|
14.5 lbs without Power Cartridge |
| Continuous Flow Setting |
0.5 - 3.0 LPM (in 0.5 liter increments) |
| Oxygen Concentration |
90% ± 3% at sea level |
| Oxygen Output Pressure |
5.0 psig (34.5 kPa) nominal |
| Sound Level |
48 dB(A) at 3.0 LPM continuous flow |
|
40 dB(A) at 3.0 pulse flow setting |
| Power Specifications |
AC Operation: 100 - 240 VAC, 50 - 60 Hz |
|
DC Operation: 12 V nominal |
|
Power Cartridge: Quantity (2) 97.5 W-Hr battery packs each containing 7.92 grams of equivalent lithium content |
| Power Consumption |
52 watts at 1.0 LPM continuous flow |
|
145 watts at 3.0 LPM continuous fl |
|
45 watts at 1.0 pulse flow setting |
|
95 watts at 6.0 pulse flow setting |
| Power Cartridge Time |
2.0 hrs @ 2.0 LPM continuous flow |
|
1.3 hrs @ 3.0 LPM continuous flow |
|
5.2 hrs @ 2.0 pulse flow setting, 12 bpm |
|
3.5 hrs @ 6.0 pulse flow setting, 12 bpm |
| Recharge Time |
1.8 - 5.0 hrs, dependent upon the flow setting, to achieve 80% capacity from a fully discharged power cartridge |
| Backup Alarm Power |
9 Volt battery |
| Oxygen Concentration Indicator |
Green light = normal operation; yellow = warning or caution; red = abnormal operation |
| Operating Temperature |
50°F to 104°F (10°C to 40°C) at 82.4°F (28°C) dew point, non-condensing |
| Pulse Flow Settings |
1.0 to 6.0 |
| Trigger Sensitivity |
Adjustable between -0.25 cm H2O to -0.7 cm H2O |
| Altitude Operating Range |
0 - 13,123 feet (0 - 4,000 meters) |
| Transition to Continuous Flow |
Yellow LED alert if no inspiration has been detected after 45 seconds; Switches to continuous flow if no inspiration has been detected for 60 seconds and a single beep |
| Alarms/Alerts |
Loss of power |
|
Low power cartridge |
|
Low therapeutic oxygen output |
|
No inspiration detected in pulse flow |
|
Unit malfunction |